Panfile parameter precise measurement and control
Panfile parameter precise measurement and control box
----A powerful tool for Panfile parameter testing
Equipment Introduction
Meets standard: IEC61215; IEC61853;
This system is used in conjunction with an IV tester to precisely control the temperature of solar modules, including a rotating mechanism. It does not affect the key indicators such as spectral level, non-uniformity, and instability of the original equipment. It can perform performance tests such as low light performance, double-sided ratio, temperature coefficient, IAM, NMOT/NOC, etc., and provide complete Panfile data.
Suitable for major battery and component factories, research and development institutions, and third-party laboratories.
Main parameters:
Thermocouple type: PT100 or T-type;
Number of thermocouples: 9;
Temperature test repeatability: ≤ 0.5 ℃;
Temperature resolution: 0.1 ℃;
Component temperature range: 15-80 ℃;
Component temperature fluctuation: ± 0.5 ℃;
Surface temperature deviation of components: ± 1 ℃;
Target temperature stabilization time: ≤ 30 minutes;
Box door opening and closing time: ≤ 2s;
Material reflectivity: ≤ 0.5%;
Effective area: 2600 * 1500mm (customizable);
IAM rotating bracket: rotation angle 0-360 °;
IAM angle resolution: 1 °;
Product advantages/core values:
1) Strong temperature control stability: Within 5 seconds of opening the door, the surface temperature fluctuation of the component is less than 0.5 ℃.
2) Good convenience: equipped with sliding rails, easy to access and place components.
3) Wide applicability: Can be used in conjunction with multiple simulators, with mobile and integrated options available.
Representing clients (including but not limited to): Canadian Solar;INTERTEK; CPTV(Luoyang); China Datang Cooperation Renewable; Changzhou Institute of Building Science.GCL System Integration Technology, etc
- File name Release date Ooperating